ONE-LINK Needle-free IV connector,
A SIMPLE ALTERNATIVE TO:

Positive Displacement at Connection

Positive Displacement at Connection

When a Positive Displacement connector is connected, blood flows back into the catheter.

Positive Displacement at Disconnection

Positive Displacement at Disconnection

When a Positive Displacement connector is disconnected, a one-time pulse of reserved fluid is pushed through the catheter tip into the bloodstream.

Neutral Displacement at Connection/Disconnection

Neutral Displacement at Connection/Disconnection

When a Neutral Displacement connector is connected or disconnected, there is minimal fluid or blood flow into or out of the catheter, helping reduce the potential for refluxing.

Negative Displacement at Disconnection

Negative Displacement at Disconnection

When proper clamping sequence is not followed, a one-time pulse of blood may flow back into the catheter. lumen upon disconnection of a Negative Displacement connector.

Neutral Displacement at Connection/Disconnection

Neutral Displacement at Connection/Disconnection

When a Neutral Displacement connector is connected or disconnected, there is minimal fluid or blood flow into or out of the catheter, helping reduce the potential for refluxing.